Table 3

Percentage of the Patellar Contact Area Subjected to Contact Pressures Greater Than 5 MPa for Normal Knees and Unresurfaced Patellae Articulating With Each Femoral Component Design

Prosthesis designKnee flexion angle
30°60°90°105°
Normal0.30.457
MG II0.353027
AMK2172628
Ortholoc0.5132222
PFC5102935
IB II0.3103431
  • MG II = Miller/Galante II prosthesis (Zimmer, Warsaw, Ind.), AMK = Anatomic Modular Knee System (DePuy, Warsaw, Ind.), Ortholoc = Whiteside Ortholoc Modular prosthesis (Wright Medical Technology, Arlington, Tenn.), PFC = press-fit condylar prosthesis (Johnson & Johnson, Warsaw, Ind.) and IB II = Insall–Burstein II prosthesis (Zimmer)