Table 1

Comparison of alignment angles achieved using FluoroGuide versus conventional surgical techniques

AngleTarget angle (range), °Group; mean (SD) [% outside of target range]Mean p value / variance95% CI
FluoroGuide (n = 53)Conventional (n = 45)
A−5 (−15 to +5)−3.8 (6.7) [6]−3.8 (6.0) [2]0.98 / 0.70−2.5 to 2.6
B0 (−5 to +10)2.8 (4.7) [11]4.2 (6.0) [18]0.20 / 0.04−0.8 to 3.6
E0 (−5 to +5)−2.3 (3.8) [15]−2.2 (5.2) [29]0.97 / 0.03−1.2 to 1.9
F7 (+2 to +12)8.0 (2.7) [6]7.3 (3.8) [4]0.30 / 0.13−2.0 to 0.6
  • CI = confidence interval; SD = standard deviation.