Abstract
Background
Use of the Harmonic Scalpel transduces a lower amount of energy to tissues, thereby limiting the potential for lateral thermal damage and deep penetration because only low temperatures are reached. The working principle of the Harmonic Scalpel is the transformation of electric power into the longitudinal mechanical movement of the instrument tip. This study aimed to determine the effects from varying durations of Harmonic Scalpel application on the experimental model of rat abdominal wall without skin.
Methods
After the rats had been anesthetized, and a laparotomy was performed. The Harmonic Scalpel shears were used on the muscular part of the abdominal wall without skin. Different durations of output power 3 were applied: a single 5-s application, a single 10-s application, and a regimen of two sequential 5-s applications. Each animal in each group of 10 received five individual activations, after which the animals were killed. Tissue samples were fixed and embedded in paraffin before sections were taken and stained. Using light microscopy and morphometric imaging analysis, the width of tissue lateral thermal damage was measured from the point of Harmonic Scalpel incision.
Results
The rat abdominal wall showed lateral thermal damage over a mean width of 0.0522 ± 0.0097 mm after a 5-s Harmonic Scalpel application, a damage width of 0.1544 ± 0.0419 mm after a 10-s application, and a damage width of 0.1020 ± 0.0430 mm after a 5-s application followed by 5 s of inactivity and another 5 s of activity. These differences in thermal damage width between all the groups are statistically significant.
Conclusions
The findings lead to the conclusion that tissue lateral thermal damage after Harmonic Scalpel application at standard output power is greater when a longer sustained period of application is used. Lateral thermal damage also is greater if the Harmonic Scalpel application time is continuous rather than of the same total duration with a brief midpoint interruption.
Similar content being viewed by others
References
Amaral JF (1995) Laparoscopic cholecystectomy in 200 consecutive patients using an ultrasonically activated scalpel. Surg Laparosc Endosc 5: 255–262
Bischof G, Zacherl J, Imhof M, Jakesz R, Fugger R (1999) Use of the Harmonic Scalpel (Ultracision) in laparoscopic antireflux surgery. Zentralbl Chir 124: 163–166
Cherqui D, Husson E, Hammoud R, Malassagne B, Stephan F, Bensaid S, Rotman N, Fagniez PL (2000) Laparoscopic liver resections: a feasibility study in 30 patients. Am Surg 66: 1037–1040
Hensman C, Baty D, Willes RG, Cuschieri A (1998) Chemical composition of smoke produced by high-frequency electrosurgery in a closed gaseous environment. Surg Endosc 12: 1017–1019
Kockerling F, Schneider C, Reck T, Scheuerlein H, Hohenberger W (1997) Laparoscopic fenestration of nonparasitic liver cysts. Zentralbl Chir 122: 1145–1149
Langer C, Markus P, Liersch T, Fuzesi L, Becker H (2001) UltraCision or high-frequency knife in transanal endoscopic microsurgery (TEM)? Advantages of a new procedure. Surg Endosc 15: 513–517
Msika S, Deroide G, Kianmanesh R, Iannelli A, Hay JM, Fingerhut A, Flamant Y (2001) Harmonic Scalpel in laparoscopic colorectal surgery. Curr Surg 58: 319–322
Ott DE, Moss E, Martinez K (1998) Aerosol exposure from an ultrasonically activated (Harmonic) device. J Am Assoc Gynecol Laparosc 5: 29–32
Perko Z, (2001) Endoscopic surgery: instruments and equipment. Knjigotisak, Split
Perko Z, Mimica Ž, Družijanić N, Kraljević D, Petričević A, Depolo A, Čala Z, Grbas H (2004) Harmonic Scalpel in laparoscopic surgery. Lijecn Vjesn 126: 246–250
Sietses C, Eijsbouts QAJ, von Blomberg BME, Cuesta MA (2001) Ultrasonic energy vs monopolar electrosurgery in laparoscopic cholecystectomy: influence on the postoperative systemic immune response. Surg Endosc 15: 69–71
Tarek A, Emam MB, Cuschieri A (2003) How safe is high-power UltraCision dissection. Ann Surg 237: 186–191
Author information
Authors and Affiliations
Rights and permissions
About this article
Cite this article
Perko, Z., Pogorelić, Z., Bilan, K. et al. Lateral thermal damage to rat abdominal wall after harmonic scalpel application. Surg Endosc 20, 322–324 (2006). https://doi.org/10.1007/s00464-005-0089-6
Received:
Accepted:
Published:
Issue Date:
DOI: https://doi.org/10.1007/s00464-005-0089-6