The importance of tibial alignment: finite element analysis of tibial malalignment

J Arthroplasty. 2000 Dec;15(8):1020-7. doi: 10.1054/arth.2000.17941.

Abstract

The influence of the tibial plateau orientation on cancellous bone stress was examined by finite element analysis for a cemented device. The objectives of the study were i) to examine the effect of the plateau-ankle angle on the cancellous bone stress, ii) to analyze the significance of the anteroposterior angles of the tibial component on these stresses, and iii) to compare the finite element predictions with clinical data. In general, positioning the tibial plateau in valgus resulted in lower cancellous bone stresses. These results support previous clinical studies, which suggest that overall alignment in valgus results in lower migration rates and lower incidence of loosening.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Biomechanical Phenomena
  • Cementation
  • Computer Simulation
  • Finite Element Analysis
  • Humans
  • In Vitro Techniques
  • Knee Prosthesis*
  • Prosthesis Failure
  • Tibia / physiology*
  • Tibia / surgery